NRIseries
Global Site> Optical Evaluation System > NRIseries  IR Refractive Index Measurement System

NRI series  IR Refractive Index Measurement System

Measurement of refractive index of infrared window, infrared lens, chalcogenide glass

 

Based on the minimum deviation method (Fraunhofer method),monochromatic light of arbitrary wavelength (1 to 14µm)is irrdiated to a prism shaped sample, and the angle of transmitted light with respect  to incident light is measured to measure the refractive index of the sample. The newly deve

 

 Catalog(PDF) <

 

 

Features

 

  • High precision measurement of the refractive index (0.0001) is possible in the wavelength range 1 to 14µm

 

  • Equipped with a temperature control function, the system can measure the difference i refractive index with temperature.

 

 

Measurement Items

 

 

Measurement Data

 

 

 

 

Block Diagram

 

 

 

Measurement Principle

 

 

 

Specifications

 

 Measurenment method 

 Monimum deviation ( Draunhofer method )
 Measurement Items

 Refractive index, Apex angle

 ・Changes in refractive index by wavelength(dn/dλ)

 ・Changes in refractive index by sample temperature(dn/dT)

 Refractive index measurement range  1~4
 Measurement precision  0.0001(1×10-4) 
 Measurement wavelength range  1~14μm
 Irrdiation wavelength range  400nm~20µm*1
 Temperature control range  -40~80℃*2
 Irrdiation size

 approx.Φ8mm

 Measurement functions  Apex angle, Determination of minimum deviation position, Refractive index
 Measurement mode  1point measurement, WL dependence, Temperature denpendence
 Setting parameters  Wavelength, Temperature, Expected refractive index
 Saving format  Text format ( CSV format )

*1 In order to perform measurement in 14 to 20µm, it is necessary to add a higher order light cut filter.
*2 Due to the set temperature of the temperature controller, it may differ depending on the measurement sample.


 

Standara Configuration

 

●Caramic heater light source
●Halogen lamp light source
●Helium lamp(d line 587.56nm)
●Light source converging system
●Monochromator and gratings
●Irrdiation optical system
●Laser for apex angle measurement
●Rotating stage and controller
●Sample holder
●Temperature control stage and controller
●Lock-in amplifer
●Sample compartment and frame
●Detector
●General controller
●Software
●Control PC

 

 

Unilities

 
●Main unit:Approx.W1800×D900×H1800mm
      *Excluding PC and rack*

 

 

Related Product

 

ART-25  Optical Elements Measuring Device

  • Mesurement repeatabilty with in ±0.2% is achieved with our proprietacy double beam optical system which cancels variations of light source and detector
  • In addition to P polarization and S polarization measurement,direct measurment of 45°N polarzed light is possible

 

KV-202 Bench Top Vacuum Ultraviolet Spectrophotometer

  • Transmission measurement in the range of 120 to300nm
  • Nitrogen purge type monocromator enables high resolution and high stabilty 