KV-202 Bench Top Vacuum Ultraviolet Spectrophometer

The KV-202 Bench top vacuum ultraviolet spectrophotometer has achieved compactness, high resolution and high stability compared to the conventional vacuum compatible type by using our original nitrogen purge type monochromator (KV-200). Light from a deuterium lamp that emits ultraviolet light is incident on a monochromator via a condensing system, is separated by a diffraction grating, and monochromatic light is extracted. The extracted monochromatic light is split into two light paths, one of which is detected as a monitor for compensating for changes in the light source, and the other is used for a semi-double beam method that irrdiates the sample, enabling highly stable transmittance spectrum measurement.

 

 Catalog(PDF)  

 

Features

 

  • Transmission measurement in the range of 120 to 300nm ( vacuum ultraviolet region )

 

  • Nitrogen purge type monochromator enables high resolution and high stability

 

  • Ideal for evaluation of special films and special substrates

 

 

Measurement Data

 

   

 

Specifictions

 

 Light source  Deuterium lamp 30W
 Wavelength range  120~300nm
 Measurement reprofucibillty  ±0.3%T *1
 Wavelength reproducibillty  ±0.1nm
 Sample size  Minimum 3×3mm ~  Maximum 12×12mm 
 Sample thickness  Minimun 1mm ~  Maximum 3mm *2
 Measurement method  Semi double beam
 Detector  PMT with a salicylic acid soda window
 Setting of display

 Scale( Y axis, X axis ), Trace ( moving cursor )

 Overlay of the spectrum, Delete of spectrum display

 Data processing

 Smoothing, Arithmetic, 1st ~ 3rd Derivatives,

 Peak processing (Peal Find Peak Height, Perk Area,Data Dump)

*1 Under the following conditions:wavelength 120~300nm, Scanning speed 60nm/min, 100% flatness
*2 Specifications of the holder shape and sample numbers to be placed can be customized as option.


 

Standard configurations

 

●Deuterium lamp 30W
●Power supply for Deuterium lamp
●KV-200 (nitrogen purge) monochromator
●Sample compartment(for nitroben purge and vacuum)
●Transmisson measurement sample holder
●PMT(photmultiplier tube) with a salicylic acid soda window
●Power supply for PMT
●Controller(wavelength drive, filter swiching) 
●Software

●PC for controlling the system

●Instruction Manual

 

 

Transmission measurement sample holder

 

 

Option

 

●Vacuum pump

●Various type of sample holder

●Various gratings

●Salicylic acid soda window

 

 

Dimensions

 

●Size : Approx.W800×D630×H500mm
●Weight : Approx.70kg
*excluding a controller and computer

 

 

Related Product

 

NRIseries  IR Refractive Index Measurement System

  • High precition measurement of the refractive index(0.0001) is possible in the wavelength range 1 to 14µm 
  • Equipped with a temperature control function,the system can measure the differrence in refrative index with temperature

 

ART-25  Optical Elements Measuring Device

  • Mesurement repeatabilty with in ±0.2% is achieved with our proprietacy double beam optical system which cancels variations of light source and detector
  • In addition to P polarization and S polarization measurement,direct measurment of 45°N polarzed light is possible