Global Site> Irrdiation System> MM-3 Multi Waveleght Irradiation System
MM-3 Multi Waveleght Irradiation System
The Model MM-3 is capable of irrdiating multiple samples simultaneously(for example, 11 pcs of rectangular cell 10mm×10mm)in each wavelength band of 220-520nm and 400-700nm.
Not only the solid samples but also liquid samples can be irradiated. Since either the irradiation time by the timer or the count numbe irradiation by the energy counter can be arbitrarily set, the total irradiation time to each sample and the total energy amount to each sample can be knouwn and the wavelength dependence test on the sample can be performed in a short time
Catalog(PDF) |
Features
- It is possble to test deterioration dependence on wavelength for various sample ( you can see at which wavelength the sample will deteriorate).
- The Model MM-3 can irradiate the sample in arbitrary wavelength range in the wavelength rage of 220~700nm.
- White light ( light which is not spectrally separated ) can also be teken out (option)
- It can irradiate any samples such as liquid samples ( drinking watar, eye drops, paint ) and solid (cosmetics, tablets, film, plastic, rubber).
Irradiation Image
Data of Light Intensity
Specficiations
Irrdiation WL range | 220~700nm WL band simultaneously irradiated is sppprox.300nm |
When 310nm is specified | 220~520nm |
When 490nm is specified | 400~700nm |
Irrdiation area | Approx.160(W)×17(H)mm |
Sample part1 | Irrdiation WL band per sample (rectangular cell 10mm) :approx.20nmWL |
Sample part2 | Interval between samples: approx.30nm ( between cell holders:15mm) |
WL display | Minimum display 1nm |
Light source |
Xe lamp 300W UV energy enhancing type reflector ari cooling |
Diffraction grating | 1200 lines/mm, 300nm blaze(MgF2 coating) |
WL dispersion | Approx.2nm/mm *The value varies depending on WL. |
Entrance slit | Width 0~10mm(continuous variable)Height 10mm(fixed) |
Exsit slit (option ) | Width 0~10mm(continuous variable) Height 10mm (fixed) |
Standard configurations
●Multi wavelength irrdiation system main unit
●Xenon lamp power supply
●Timer controller
●Glass filter set for 310nm
●Glass filter set for 490nm
●Sample mask
●Quartz glass cell 10mm×10mm(11pcs )
●Quarts cell holder ( for 11 pcs cells )
●Ozon exhaust hose( 5m )
●Instruction Manual
Options
●Monochromatic light extraction mechanism
●White light extraction mechanism
●Power meter for irrdiation energy
●Glass filter set ( need to spectify WL )
●Sample holder for various samples
Light path figure
Dimensions
●Main unit :approx.W630×D620×H370mm
●Power supply :approx.W175×D410×H225mm
●Controller :approx.W215×D315×H110mm