MM-3

Global Site> Irrdiation System> MM-3 Multi Waveleght Irradiation System 

MM-3 Multi Waveleght Irradiation System

The Model MM-3 is capable of irrdiating multiple samples simultaneously(for example, 11 pcs of rectangular cell 10mm×10mm)in each wavelength band of 220-520nm and 400-700nm.

Not only the solid samples but also liquid samples can be irradiated. Since either the irradiation time by the timer or the count numbe irradiation by the energy counter can be arbitrarily set, the total irradiation time to each sample and the total energy amount to each sample can be knouwn and the wavelength dependence test on the sample can be performed in a short time

 Catalog(PDF)

 

 

Features

 

  • It is possble to test deterioration dependence on wavelength for various sample ( you can see at which wavelength the sample will deteriorate).

 

  • The Model MM-3 can irradiate the sample in arbitrary wavelength range in the wavelength rage of 220~700nm.

 

  • White light ( light which is not spectrally separated ) can  also be teken out (option)

 

  • It can irradiate any samples such as liquid samples ( drinking watar, eye drops, paint ) and solid (cosmetics, tablets, film, plastic, rubber).

 

 

Irradiation Image

 

 

Data of Light Intensity

 

 

 

Specficiations

 

 Irrdiation WL range  220~700nm WL band simultaneously irradiated is sppprox.300nm
 When 310nm is specified  220~520nm
 When 490nm is specified  400~700nm
 Irrdiation area  Approx.160(W)×17(H)mm
 Sample part1  Irrdiation WL band per sample (rectangular cell 10mm) :approx.20nmWL
 Sample part2  Interval between samples: approx.30nm ( between cell holders:15mm)
 WL display  Minimum display 1nm
 Light source

 Xe lamp 300W  UV energy enhancing type reflector ari cooling

 Diffraction grating  1200 lines/mm, 300nm blaze(MgF2 coating)
 WL dispersion  Approx.2nm/mm *The value varies depending on WL. 
 Entrance slit  Width 0~10mm(continuous variable)Height 10mm(fixed)
 Exsit slit (option )  Width 0~10mm(continuous variable)  Height 10mm (fixed)

 

 

Standard configurations

 

●Multi wavelength irrdiation system main unit

●Xenon lamp power supply

●Timer controller

●Glass filter set for 310nm

●Glass filter set for 490nm

●Sample mask
●Quartz glass cell 10mm×10mm(11pcs )

●Quarts cell holder ( for 11 pcs cells )

●Ozon exhaust hose( 5m )

●Instruction Manual

 

 

Options


●Monochromatic light extraction mechanism
●White light extraction mechanism
●Power meter for irrdiation energy
●Glass filter set ( need to spectify WL )
●Sample holder for various samples

 

 

Light path figure

 

 

Dimensions

 

●Main unit       :approx.W630×D620×H370mm
●Power supply :approx.W175×D410×H225mm
●Controller      :approx.W215×D315×H110mm

 

 

Related Product

 
NIJI-2 Variable Wavelength Light Source

 
  • Irradiation in the wide range from ultraviolet,visble,to Infrared light can becoverd
  • Stand-alone and compact desigh require only a modest requirement of bench space for installation 

 

IUV-25CP  Spectral Irradiation System For Photoresist

 
  • Testing the spectral sensitivity of photoresist agents can be performed
  • The travel distance and the nunber of steps can be set to 5mm×13 stpes at maximum

 

BIR-50  Solar Cell Light Resistance Testing System

 
  • Long accurate light degradation test with the solar simulator
  • I-V measurement under constant temperature atmosphere(20~90℃)