BIR-50
Global Site> Irrdiation System> BIR-50 Solar Cell Light Resistance Testing System 

BIR-50 Solar Cell Light Resistance Testing System

 

Since the conventional light test system is not a dedicated system for testing solar cells, the accuracy of the irrdiation light is not enough when using a xenon lamp or a metal hailde lamp or etc., and there have beeen three problems as follows.


1.Effect of the spectral mismatch( influence of ultraviolet deterioration and infrared deterioration due to the intensity difference between the sunlight and irrdiation light )


2.Effect of non-uniformity of irrdiation light ( influence of variation of irradiation intensity on the individual sample when setting the multiple samples, )

3.Effect of irrdiation time ( influence of the change of irradiation light intensity when starting measurement and completing measurement ). 

Therefore, the Model BIR-50 has been developed to combine the conventional solar simulator with 50mm×50mm irrdiation area ( class AAA ) and constant temperature chamber for the temperture adjustment, which is capable to perform more accurate light degradation test and the temperature deterioration test of the solar cell. There is the BNC connetor equipped in the constant temperature chamber that is enabled to perfrom the I-V measurement.

 

 Catalog(PDF)

 

Features

 

 

  • Long accurate light degadiatin test with the solar simulator (class AAA)

 

  • Excellent stabilty of light intensity such as within±1% with our unique light intensity intensity feedback system

 

  • I-V measurement under constant temperature atmosphere (20~90℃)

 

  • Degradation test with specific wavelength range setting the-optional filters

 

 

Light intensity  feedback with an internal monitor detetor

An internal monitor detector is bulit in the optical system.

Light intensity dan be maintained constant by monitoring the irrdiation light intensity and feeding it back to the Xenon light power supply.

 

 

I-V measurement under constant temperature

The BNC connector are equipped with in and outside the constant temperature chamber. I-V measurement can be performed under constant temperature.

 

 

 

Observation of the sample during irrdiating the light on the sample

 The temperature cannot be controlled when the door is opened during irradiating the light on the sample in order to observe the sample. Therefore, the BIR-50 has been developed with the observation window, which enables the customer to observe the sample without opening the door while maintaining the setting temperature in the constant temperature chamber.

 

 

Equipped with a filter holder

 

Equipped with a filter holder in the optical system, you can use up to four filters at the same time.

Degradation when the light is attenuated with the ND filters or Degradiation in different wavelength range with different color filters can be performed. ND filters and color filters and filters are optional items.

 

 

Specifications

 

 Irrdiation Intensity  1SUN ( equivalent to 100mW/cm2 )
 Irrdiation area  50×50mm
 Spectral match  Class A  ( 0.75~1.25 )   *JIS C8912 C8933
 Non-uniformity  Class A  (with in ±2% )  *JIS C8912 C8933
 Instability of irradiance  Class A  (with in ±1% )*JIS C8912 C8933
 Setting temperature range  20~90℃
 Temperature precision  Within ±1℃(within ±0.5 between 25~60℃)
 Humidity measurement range

 0~100%RH *Humidity can not controlled*

 Humidity precision

 ±2%(@25℃ 0~90%RH)

 

 

Option

 

●Color filter
●ND filter

●Various stage for back electrode type sample, DSC sample stage and etc.

 

 

Dimensions


●Approx. : W1000×D620×H700mm

 *Xenon power supply, ammeter and PC are not included.

 

Related Product

 
NIJI-2 Variable Wavelength Light Source

 
  • Irradiation in the wide range from ultraviolet,visble,to Infrared light can becoverd
  • Stand-alone and compact desigh require only a modest requirement of bench space for installation 

 

MM-3  Multi Waveleght Irradiation System

 
  • It is possible to test deterioration dependence on wavelenth for various samples
  • The Model MM-3 can irradiate the sample in arbitrary wavelenght range in the wavelength rage of 220~770nm

 

IUV-25CP  Spectral Irradiation System For Photoresist

 
  • Testing the spectral sensitivity of photoresist agents can be performed
  • The travel distance and the nunber of steps can be set to 5mm×13 stpes at maximum